FLOW CHART OF THE STERILIZATION PROCEDURE.

deposit the
product and
close the door

heating

evacuation injecting
the E.O Gas
sterilization
during 8 hours
               
End of
sterilization
process
inject
the air
and open
evacuation discharging
the E.O Gas
stop heating


STERILIZATION IN DETAIL

1.
2.
3.
4.
5.
6.
7.
8.
9.
Put the products inspected and packed to the sterilization device.
Heating.
Make the chamber vacuous.
Inject the E.O Gas to the chamber.
Sterilize for 8 hours.
Turn off the heating.
Discharge the E.O. Gas.
Make the chamber vacuous.
Supply air and take out the products


THE CONDITIONS OF STERILIZATION.

1.
2.
3.
4.
5.
Temperature of the chamber : 55¡É
Humidity of the chamber : 68%
Pressure of the vacuum : below than 68.9kpa
Pressure of the gas : higher than 1§¸/§¨
The period of the sterilization : more than 8 hours


STERILIZER OF E.O GAS

 

 

 